产品中心 药物筛选试剂盒 UniOne®TR-FRET

UniOne® TR-FRET Human BRD4(BD1)/H4 Binding Kit

价格 4,380.00 1-2周
货号 UA086143
规格
数量
收藏 分享

产品介绍 评论(0)

背景介绍
  • ASSAY PRINCIPLE

    The kit is based on homogeneous time-resolved fluorescence technology (TR-FRET) technology,which is used to assess the molecular-mediated interaction between the BRD4 (BD1) and the H4 peptide. This method provides a high-throughput approach to detect the small molecules capable of mediating the interaction between BRD4 (BD1) and H4 peptide simply and rapidly.

    As shown in the figure below, the interaction between BRD4 (BD1) and H4 peptide was detected by an Eu-labeled anti-Tag1 antibody (TR-FRET donor) and an Ac-labeled anti-Tag2 antibody (TR-FRET receptor). The donor is in close proximity to the receptor antibody,excitation of the donor antibody triggers fluorescence resonance energy transfer (FRET) to the receptor antibody, resulting in specific emission of a signal at a wavelength of 665 nm from the receptor antibody. Since the molecular (+) JQ-1 block the interaction between BRD4 (BD1) and H4 peptide, the FRET signal is reduced. This specific signal is proportional to the degree of interaction between BRD4(BD1)and H4 peptide. This homogeneous assay is simple to perform and requires no washing steps.

产品规格
  • 来源

    Human
  • 储存条件

    -80℃

产品组分
  • 【Composition and Storage Conditions】

    Composition

    concentration

    100T

    500T

    2500T

    10000T

    Storage

    Tag1-BRD4 (BD1)

    100×

    5μL

    20μL

    100μL

    400μL

    -80℃

    Tag2-H4 peptide

    100×

    5μL

    20μL

    100μL

    400μL

    -80℃

    Anti-Tag1 Eu antibody

    10

    5μL

    25μL

    125μL

    500μL

    -80℃

    Anti-Tag2 Ac antibody

    12.5×

    40μL

    200μL

    1mL

    4mL

    -80℃

    Detection buffer

    4mL

    20mL

    100mL

    400mL

    -80℃


    Note: Aliquot into appropriate volumes, store at the recommended temperature and avoid repeated freeze-thaw cycle.

操作步骤
  • 【Process and Operations】

    1、Reagent Preparation

    1.1 Compositions equilibration

    Before use, dissolve all reagents at room temperature (allow at least 30 minutes for temperature equilibration). For the 384-well plate assay, use 20 μL of reaction mixture (the reagent quantities are listed in the table below). Calculate the required volume prior to preparation accordingly. The following preparation steps are provided for reference only, using a 500-well plate as an example.

    Table 1. Reagent Preparation

    Composition

    Configuration

    μL/well

    Tag1-BRD4 (BD1)

    Take 20 μL of the 100× Tag1-BRD4 (BD1) stock solution, add 1.98 mL of Detection buffer to dilute it to 1× concentration, and mix thoroughly for later use.

    4

    Tag2-H4 peptide

    Take 20 μL of 100× Tag2-H4 peptide, add it to 1.98 mL of Detection buffer to dilute to 1× concentration, and mix thoroughly for later use.

    4

    Antibody Mix

    Take 25 μL of the 100× Anti-Tag1 Eu antibody stock solution, add 2475 μL of Detection buffer to dilute to 2.5 mL, and mix thoroughly; take 200 μL of the 12.5× Anti-Tag2 Ac antibody stock solution, add 2300 μL of Detection buffer to dilute to 2.5 mL, and mix thoroughly; combine the two solutions in a 1:1 ratio to obtain the Antibody Mix.

    10


    1.2 Gradient dilution of the sample

    (+) JQ-1 as an example, the dilution solution is Detection buffer. To minimize interference from matrix effects, it is recommended to use a solution with the same matrix composition as the sample being tested; additionally, the sample concentration should be adjusted according to its actual situation.

    Table 2. (+) JQ-1 gradient dilution(Adjust according to the actual situation)

    (+) JQ-1 final concentration(nM)

    (+) JQ-1 preparation concentration(nM)

    Dilution

    300

    3000

    0.5μL 500 μM (+)JQ1 + 82.85μL Detection buffer

    100.00

    1000.00

    10μL ① +20μL Detection buffer

    33.33

    333.33

    10μL ② +20μL Detection buffer

    11.11

    111.11

    10μL ③ +20μL Detection buffer

    3.70

    37.04

    10μL ④ +20μL Detection buffer

    1.23

    12.35

    10μL ⑤ +20μL Detection buffer

    0.41

    4.12

    10μL ⑥ +20μL Detection buffer

    0.14

    1.37

    10μL ⑦+20μL Detection buffer

    0.05

    0.46

    10μL ⑧+20μL Detection buffer

    0.02

    0.15

    10μL ⑨+20μL Detection buffer

    Blank

    0

    0

    20μL Detection buffer


    2、Sample and control

    2.1 Sample:Add 4 μL of Tag1-BRD4 (BD1) working solution, 4 μL of Tag2-H4 peptide working solution, 2 μL of the gradient-diluted sample , and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

    2.2 (+) JQ-1 : 4 μL of Tag1-BRD4 (BD1) working solution, 4 μL of Tag2-H4 peptide working solution, 2 μL of gradient-diluted (+) JQ-1 , and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

    2.3 Blank:4 μL of Tag1-BRD4 (BD1) working solution, 4 μL of Tag2-H4 peptide working solution,, 2 μL of Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

    2.3 Negative control:10μL Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

    After adding all samples, centrifuge, seal with a cover membrane, and incubate at room temperature for 2 hours.

    Table 3. Sample and control spotting process

    (+) JQ-1

    Sample

    Blank

    Negative control (NC)

    2μL

    gradient-diluted (+) JQ-1

    2μL

    gradient-diluted sample

    2μL

    Detection buffer

    10μL Detection buffer

    and 10μL Antibody Mix

    4μL Tag1-BRD4 (BD1)

    Incubate at room temperature for 10min

    4μL Tag2-H4 peptide

    10μL Antibody Mix

    Seal the plate wells with a sealing film and incubate at room temperature for 2 hours.


    3、Read Data

    The detection was performed on a TR-FRET microplate reader. The excitation wavelength was 320/340 nm, and the emission wavelengths at 620 nm and 665 nm were detected.

    【DATA REDUCTION 】

    1) Ratio:

    Ratio = (665/620) ×10000

    2) Net signal:

    Net signal = (Std-NC)/NC×100

    Std:the ratio of sample and control

    NC: the ratio of negative control

    3) CV(%):

    CV(%)= Standard Deviation/Mean Ratio × 100%

    【Data Example】

    The following data cannot replace experimental results obtained in experiments; they are provided solely for illustrative purposes, and outcomes may vary depending on the plate reader used.

    Note: Recommended microplate (384-well plate, white, shallow wells).

评论(0)