【Process and Operations】
1、Reagent Preparation
1.1 Compositions equilibration
Before use, dissolve all reagents at room temperature (allow at least 30 minutes for temperature equilibration). For the 384-well plate assay, use 20 μL of reaction mixture (the reagent quantities are listed in the table below). Calculate the required volume prior to preparation accordingly. The following preparation steps are provided for reference only, using a 500-well plate as an example.
Table 1. Reagent Preparation
Composition | 配置 | μL/well |
Tag1-CRBN protein | Take 20 μL of the 100× Tag1-CRBN protein stock solution, add 1.98 mL of Detection buffer to dilute it to 1× concentration, and mix thoroughly for later use. | 4 |
Tag2-CDK4 protein | Take 16 μL of 126× Tag2-CDK4 protein stock solution, add it to 2 mL of Detection buffer to dilute to 1× concentration, and mix thoroughly for later use. | 4 |
Antibody Mix | Take 50 μL of the 50× Anti-Tag1 Eu antibody stock solution, add 2450 μL of Detection buffer to dilute to 2.5 mL, and mix thoroughly; take 25 μL of the 100×Anti-Tag2 Ac antibody stock solution, add 2475 μL μL of Detection buffer to dilute to 2.5 mL, and mix thoroughly; combine the two solutions in a 1:1 ratio to obtain the Antibody Mix. | 10 |
1.2 Gradient dilution of the sample
BSJ-03-204 as an example, the dilution solution is Detection buffer. To minimize interference from matrix effects, it is recommended to use a solution with the same matrix composition as the sample being tested; additionally, the sample concentration should be adjusted according to its actual situation.
Table 2. BSJ-03-204 gradient dilution(Adjust according to the actual situation)
| BSJ-03-204 final concentration(nM) | BSJ-03-204 preparation concentration(nM) | Dilution |
① | 2000 | 20000 | 2μL 1mM BSJ-03-204 +98μL 1× Detection buffer |
② | 1000 | 10000 | 30μL ① +30μL Detection buffer |
③ | 500 | 5000 | 30μL ② +30μL Detection buffer |
④ | 250 | 2500 | 30μL ③ +30μL Detection buffer |
⑤ | 125 | 1250 | 30μL ④ +30μL Detection buffer |
⑥ | 62.5 | 625 | 30μL ⑤ +30μL Detection buffer |
⑦ | 31.25 | 312.5 | 30μL ⑥ +30μL Detection buffer |
⑧ | 15.625 | 156.25 | 30μL ⑦+30μL Detection buffer |
⑨ | 7.8125 | 78.125 | 30μL ⑧+30μL Detection buffer |
⑩ | 3.90625 | 39.0625 | 30μL ⑨+30μL Detection buffer |
Blank | 0 | 0 | 30μL Detection buffer |
2、Sample and control
2.1 Sample:Add 4 μL of Tag1-CRBN protein working solution, 4 μL of Tag2-CDK4 protein working solution, 2 μL of the gradient-diluted sample , and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.
2.2 BSJ-03-204: 4 μL of Tag1-CRBN protein working solution, 4 μL of Tag2-CDK4 protein working solution, 2 μL of gradient-diluted BSJ-03-204, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.
2.3 Blank:4 μL of Tag1-CRBN protein working solution, 4 μL of Tag2-CDK4 protein working solution, 2 μL of Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.
2.3 Negative control: 10μL Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.
After adding all samples, centrifuge, seal with a cover membrane, and incubate at room temperature for 2 hours.
Table 3. Sample and control spotting process
BSJ-03-204 | Sample | Blank | Negative control(NC) |
2μL gradient-diluted BSJ-03-204 | 2μL gradient-diluted sample | 2μL Detection buffer | 10μL Detection buffer and 10μL Antibody Mix |
4μL Tag1-CRBN protein |
4μL Tag2-CDK4 protein |
10μL Antibody Mix |
Seal the plate wells with a sealing film and incubate at room temperature more than 2 hours. |
3、Read Data
The detection was performed on a TR-FRET microplate reader. The excitation wavelength was 320/340 nm, and the emission wavelengths at 620 nm and 665 nm were detected.
【DATA REDUCTION 】

1) Ratio:Ratio = (665/620) ×10000
2) Net signal:
Net signal = (Std-NC)/NC×100
Std:the ratio of sample and control
NC: the ratio of negative control
3) CV(%):
CV(%)= Standard Deviation/Mean Ratio × 100%
【Data Example】
The following data cannot replace experimental results obtained in experiments; they are provided solely for illustrative purposes, and outcomes may vary depending on the plate reader used.
Note: Recommended microplate (384-well plate, white, shallow wells).